top of page
LOGO2

 

DROP DOWN MENU

Trace of 1mM addns to 0.1% NaCl
 Calibration with NaCl Additions
1mM additions of NaCl to 0.1% NaCl

Temperature Control & Stability

Temperature Warmup & Stability

Simple Test using NaCl to demonstrate linearity of                  

     apparatus response to changes in [NaCl]

The graph above refers to the circuit which used the long-tailed pair differential amplifier. The chart below shows the equivalent result using the 741 opamp amplifier

The graph below shows the original data used to construct the chart above. Note that the sensitivity of the 741 circuit is approximately half that of the long-tailed pair. However as mentioned elesewhere the noise is considerable better owing to common mode noise reduction in the input square wave.

 

It is obvious that 10% of this addition i.e. 100 micromolar can easily be measured and 1% i.e. 10 micromolar is possible with attention to detail.

CALIBRATION

Double click to insert body text here ...

The volume given above

is incorrect it should

   read 0.0067 ml

Trace of 1mM addns to 1% NaCl

The graph below shows the same 1 mM additions to 1% NaCl. This is much more difficult to measure against a strong ionic background. The measurement is also less sensitive

 

The cell constants were measured as 1.27 & 1.34 cm-1 for the two cells

   The calculated value is 1.2, taking fringe fields into account

 

 

Impedance of the electrode assemblies

For optimum sensitivity of a Wheatstone bridge the four resistances should be approximately the same at the start of an experiment. For an AC bridge the reactances (or impedances) should be similar.

 

The reactance of the buffer used in enzyme experiments is ca. 500 ohms for the 28 kHz square wave used in the bridge. [This increases during a 36 microsecond period of the half wave if polarisation occurs]. The reactances of the two cells are 670 & 633 ohms for a 0.1% NaCl solution that was used to calibrate & test the system.

 

This is the reason that a 1 k ohm potentiometer was chosen to balance the bridge, since each half gives a resistance of 500 ohms at balance before the start of an experiment & this leads to optimum sensitivity of the bridge. It is advisable to change the resistance of the potentiometer if a very different buffer is used to give as good a match as is practicable. However tests have shown that reasonable results can be obtained even if the buffer strength is reduced 10 – fold, while the potentiometer is unchanged (although clearly 10 k ohm pot would be better). For best results at lowest concentrations change the pot!

 

More generally a change of 1% in the overall ionic strength can be measured when the pot is approximately (+/- 50%) matched to the impedance of the reaction solution. If the pot rsistance differs by more than 10-fold from the solution impedance the best that can be measured is a 10% change. This is a general rule that is valuable in planning experiments.

 

 

 

 

 

 

Graph of 1mM addns to 1% NaCl
LOG-LOG Plot of apparatus Sensitvity with [NaCl]
Table of Apparatus Sensitivity with [NaCl]

The Table below shows how the sensitivity of the apparatus changes with the salt concentration. The Log-Log plot below the Table shows that the sensitivity is nearly inversely proportional to the salt concentration. This empahsises how important it is to choose an appropriate buffer concentration for your enzyme assays. You do not want the pH to change during the assay but also you want to minimise the salt concentration, i.e. the buffer concentration. The buffer concentration should be a minimum of 10 times the change in ions measured as a result of the enzyme reaction in the assay.

 

MAKE SURE THAT ADDITION OF THE ENZYME SOLUTION DOES NOT CHANGE THE pH - DIALYSE AGAINST REACTION BUFFER IF NECESSARY TO AVOID THIS.

Temperature Control                   and Calibration

 

Temperature Resilience

NaCl Tests

Urea/Urease Test

WEB 1 micromolar addns to 0-0001% NaCl

1 micromolar additions to 0.0001% NaCl [16.7 micromolar].

This trace shows how good the sensitivity is at low    background ionic concentrations

    Some general Rules

 

Tests with Sodium Chloride

 

   Test with urease & Urea

UREA/urease SENSITIVITY TEST

In the second of the references quoted in the ref section it is reported that the hydrolysis of 2 nmoles of urea could be detected using the precision apparatus used for these experiments. I thus set out to try & demonstrate that the apparatus used here could detect a 1 nM change. In the event I was stymied by a high salt concentration (400 mM!) in the enzyme stock solution (100 mg/ml crude jack bean powder), which I was unable to remove as I do not have enough material for dialysis. The reaction solution (after dilution of 0.01 ml enzyme to 1 ml) was 4 mM in salt, too high to measure a 1 nmole change.

 

I thus tried to detect hydrolysis of 10 nmoles of urea, with success. The graph below shows two 10 nmole additions.

UREA/UREASE Sensitivity TEST

Response to 1 degree temperatrure change at balance
Response to 1 degree temperature change away from balance

Temperature Resilience

 

The graphs below shows the temperature resilience that comes from using two active electrochemical cells in the bridge rather than just one. The left hand side graph shows the response to a 1 degree rise in temperature within the box at balance with 0.1% NaCl in both the cells. There is little change because the effects in each cell cancel each other.

The right hand side graph above shows the response to a 1 degree temperature rise when the instrument is away from balance. 5 mM NaCl was added to the sample cell, giving a voltage of ca. 1V. Note that the voltage change is now larger as would be expected.

bottom of page